Doppler Optical Coherence Tomography of Retinal Circulation
Published on: September 18, 2012
Are quantitative attenuation measurements of blood by optical coherence tomography feasible?
Dirk J Faber1, Ton G van Leeuwen
1Ophthalmology Department, Academic Medical Center, University of Amsterdam, PO Box 22700, 1100 DE Amsterdam, the Netherlands. d.j.faber@amc.uva.nl
Abstract:
We present optical coherence tomography (OCT) measurements on fully physiologically oxygenated blood samples of varying hematocrit. We show that attenuation coefficients cannot be extracted quantitatively using the currently accepted models for the OCT signal from scattering media, because the confidence intervals obtained in the fitting procedure cannot be used as reliable uncertainty estimates of the attenuation coefficients. Better modeling of the hematocrit-dependent OCT signal is needed.