Quantitation of human immunodeficiency virus type 1 viral load in plasma using reverse transcriptase activity assay

Madisa Mine1, Keabetswe Bedi, Talkmore Maruta

  • 1National Health Laboratory, 5353 Church Road, Extension 10, Gaborone, Botswana. mmine@bhp.org.bw

Insights

The Cavidi ExaVir Load v2 assay offers a reliable, cost-effective alternative for HIV viral load monitoring in Botswana. This study found it comparable to the gold standard COBAS Amplicor monitor v1.5, enabling decentralized testing.

Area of Science:

  • Virology
  • Clinical Diagnostics
  • Public Health

Background:

  • The Roche COBAS Amplicor monitor v1.5 assay is the gold standard for HIV viral load monitoring in Botswana.
  • Its infrastructure requirements limit testing to national reference laboratories, hindering access in rural areas.

Purpose of the Study:

  • To evaluate the Cavidi ExaVir Load v2 assay as a decentralized viral load testing alternative.
  • To compare its performance against the COBAS Amplicor monitor v1.5 assay in a district hospital setting.

Main Methods:

  • Performance comparison of ExaVir Load v2 and COBAS Amplicor monitor v1.5 assays.
  • Utilized quality assessment samples and plasma from HIV-positive individuals.
  • Statistical analysis included Kappa statistic and correlation coefficient.

Main Results:

  • High agreement between ExaVir Load v2 and COBAS Amplicor monitor v1.5 assays (Kappa = 0.951).
  • Strong correlation observed (r=0.8554, P<0.0001).
  • ExaVir Load v2 detected HIV-1 RNA in 86 samples, COBAS in 84; concordant detection in 82 samples.

Conclusions:

  • Cavidi ExaVir Load v2 assay is a simple, reliable, and adaptable alternative for decentralized viral load testing.
  • It offers a lower cost per test compared to RT-PCR.
  • ExaVir Load v2 systems have been implemented in multiple district and primary hospitals.