Scan-time reduction using noise-matched images in 2- and 3-dimensional bismuth germanate PET/CT: clinical study in
Bal Sanghera1, Gerry Lowe, David Wellstead
1Paul Strickland Scanner Centre, Mount Vernon Hospital, Middlesex, United Kingdom. bal.sanghera@nhs.net
Unlabelled:
We quantitatively and qualitatively investigated 2-dimensional (2D) and 3-dimensional (3D) imaging with scan-time reduction in 14 patients with 17 lesions, who had known or suspected head and neck cancer, using a bismuth germanate (BGO) crystal based PET/CT scanner with noise-matched images.
Methods:
A 2D and 3D acquisition protocol using scan-time reduction on an axial single field of view resulted in a 2D 4-, 3D 4-, 3D 3-, 2D 3-, 2D 2-, and 3D 2-min scan sequence to minimize redistribution and decay bias. Tumor maximum standardized uptake values (SUVmax) and tumor mean standardized uptake values (SUV(mean)) were recorded, and two observers in consensus investigated lesion conspicuity between 2D and 3D paired 4-, 3-, and 2-min noise-matched images.
Results:
We found some minor advantages quantitatively in favor of 2D scanning, with higher mean SUVmax, and qualitatively in favor of 3D scanning, with lesion conspicuity preference. In our cohort, no great advantage or disadvantage to using either acquisition mode was observed, and all lesions were seen irrespective of acquisition mode and scan time.
Conclusion:
In head and neck cancer patients, we can recommend a scan-time reduction from 4 to 3 min/bed position in 2D acquisitions with a BGO-based PET/CT scanner, using our imaging protocol and reconstruction defaults.
More Related Videos
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET