Related Experiment Video
Updated: Jun 23, 2026

Trabecular Bone Microarchitecture Evaluation in an Osteoporosis Mouse Model
Published on: September 8, 2023
Assessment of trabecular and cortical architecture and mechanical competence of bone by high-resolution peripheral
A Cohen1, D W Dempster, R Müller
1Department of Medicine, PH8-864, Columbia University, College of Physicians and Surgeons, 630 West 168th Street, New York, NY 10032, USA. ac1044@columbia.edu
Unlabelled:
We compared microarchitecture and mechanical competence parameters measured by high-resolution peripheral quantitative computed tomography (HR-pQCT) and finite-element analysis of radius and tibia to those measured by histomorphometry, micro-CT, and finite-element analysis of transiliac bone biopsies. Correlations were weak to moderate between parameters measured on biopsies and scans.
Introduction:
HR-pQCT is a new imaging technique that assesses trabecular and cortical bone microarchitecture of the radius and tibia in vivo. The purpose of this study was to determine the extent to which microarchitectural variables measured by HR-pQCT reflect those measured by the "gold standard," transiliac bone biopsy.
Methods:
HR-pQCT scans (Xtreme CT, Scanco Medical AG) and iliac crest bone biopsies were performed in 54 subjects (aged 39 +/- 10 years). Biopsies were analyzed by 2D quantitative histomorphometry and 3D microcomputed tomography (microCT). Apparent Young's modulus, an estimate of mechanical competence or strength, was determined by micro-finite-element analysis (microFE) of biopsy microCT and HR-pQCT images.
Results:
The strongest correlations observed were between trabecular parameters (bone volume fraction, number, separation) measured by microCT of biopsies and HR-pQCT of the radius (R 0.365-0.522; P < 0.01). Cortical width of biopsies correlated with cortical thickness by HR-pQCT, but only at the tibia (R = 0.360, P < 0.01). Apparent Young's modulus calculated by microFE of biopsies correlated with that calculated for both radius (R = 0.442; P < 0.001) and tibia (R = 0.380; P < 0.001) HR-pQCT scans.
Conclusions:
The associations between peripheral (HR-pQCT) and axial (transiliac biopsy) measures of microarchitecture and estimated mechanical competence are significant but modest.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...