Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Experiment Video

Updated: Jun 23, 2026

Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model
05:52

Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model

Published on: November 4, 2021

Microcoil embolization through a downsized coaxial catheter system: an experimental study.

T Tajima1, K Yoshimitsu, H Irie

  • 1Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan. ttajima@med.kyushu-u.ac.jp

Acta Radiologica (Stockholm, Sweden : 1987)
|May 21, 2009
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Prediction of pathological response to preoperative chemotherapy for pancreatic ductal adenocarcinoma using 2-[<sup>18</sup>F]-fluoro-2-deoxy-d-glucose positron-emission tomography.

Clinical radiology·2022
Same author

Image quality and radiation dose of renal perfusion CT with low-dose contrast agent: a comparison with conventional CT using a 320-row system.

Clinical radiology·2019
Same author

Characterization of Pathogenic Races of Fusarium oxysporum f. sp. melonis Causing Fusarium Wilt of Melon in New York.

Plant disease·2019
Same author

Clinical Outcomes of Living Liver Transplantation According to the Presence of Sarcopenia as Defined by Skeletal Muscle Mass, Hand Grip, and Gait Speed.

Transplantation proceedings·2017
Same author

Predictors of therapeutic effect of transarterial chemoembolisation using drug-eluting beads for hepatocellular carcinoma.

Clinical radiology·2017
Same author

Fatty change in moderately and poorly differentiated hepatocellular carcinoma on MRI: a possible mechanism related to decreased arterial flow.

Clinical radiology·2016

This study shows that microcoil embolization is feasible using a new 2-F microcatheter (MC6) with a 3-F system. All tested microcoils passed successfully, demonstrating compatibility for abdominal angiography procedures.

Area of Science:

  • Interventional Radiology
  • Vascular Surgery
  • Medical Device Engineering

Background:

  • Existing microcatheters often have large trailing end diameters, limiting their use in 3-French (F) systems for microcoil embolization.
  • This incompatibility hinders minimally invasive treatment options in abdominal angiography.

Purpose of the Study:

  • To evaluate the in vitro passage of various microcoils through a novel 2-F microcatheter (Meister Cath Superselective Plus, MC6).
  • To assess the compatibility of MC6 with a downsized coaxial 3-F catheter system for embolization procedures.

Main Methods:

  • In vitro testing of microcoil passage through the MC6 microcatheter within a blood vessel model.
  • Utilized six types of microcoils in 17 size variations (2-10 mm diameter, 20-140 mm length).

More Related Videos

Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
11:27

Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging

Published on: April 4, 2013

Microvascular Embolism Mouse Model for In Vivo Two-photon Microscopy Using Fluorescent Polystyrene Microspheres
08:29

Microvascular Embolism Mouse Model for In Vivo Two-photon Microscopy Using Fluorescent Polystyrene Microspheres

Published on: November 21, 2025

Related Experiment Videos

Last Updated: Jun 23, 2026

Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model
05:52

Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model

Published on: November 4, 2021

Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
11:27

Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging

Published on: April 4, 2013

Microvascular Embolism Mouse Model for In Vivo Two-photon Microscopy Using Fluorescent Polystyrene Microspheres
08:29

Microvascular Embolism Mouse Model for In Vivo Two-photon Microscopy Using Fluorescent Polystyrene Microspheres

Published on: November 21, 2025

  • Assessed microcoil passage, post-deployment shape, and saline volume required for flushing.
  • Main Results:

    • All tested microcoils successfully passed through the MC6 without catheter occlusion.
    • Satisfactory deployment was achieved for all microcoils.
    • Significant differences in saline flush volumes were noted between certain microcoil types (e.g., VortX vs. Reverse TORNADO).

    Conclusions:

    • Microcoil embolization using the new 2-F MC6 microcatheter within a 3-F system is feasible.
    • The MC6 demonstrates compatibility with various microcoils for potential use in abdominal angiography.