Effects of AEC chamber selection on patient dose and image quality
1University of Arkansas-Fort Smith, AR, USA.
Purpose:
To determine whether manipulation of the standard automatic exposure control (AEC) chamber selections reduces the patient's entrance skin exposure (ESE) without compromising image quality.
Methods:
Data for density and radiation dose were gathered at 2 clinical locations by exposing abdomen and pelvis phantoms to radiation using 3 AEC chamber selection configurations. ESE (skin dose) was measured using a multipurpose dosimeter. The experiment included both film-screen and computed radiography (CR) systems.
Results:
For both phantoms, using the 2 outside chambers resulted in the lowest dose on the film-screen and CR systems. In general, optical density (OD) and exposure indicator (EI) remained within acceptable ranges and image quality was maintained using this chamber configuration. Using only the center chamber resulted in the highest dose increases and lowest image quality for film-screen and CR systems.
Conclusion:
When performing anteroposterior (AP) abdomen and AP pelvis examinations, radiographers can reduce patients' ESE and maintain image quality by selecting the 2 outside AEC chambers. Further research on AEC chamber selection should be conducted for additional anatomical regions.