Desflurane consumption with the Zeus during automated closed circuit versus low flow anesthesia
S De Cooman1, A Lecain, M Sosnowski
1Department of Anesthesiology, Institut Jules Bordet, Université Libre de Bruxelles, Brussels, Belgium.
Introduction:
During automated closed-circuit anesthesia (CCA), the Zeus (Dräger, Lübeck, Germany) uses a high initial fresh gas flow (FGF) to rapidly attain the desired agent and carrier gas concentrations, resulting in a desflurane consumption well above patient uptake. Because both FGF and carrier gas composition can affect consumption, we determined the Zeus' agent consumption with automated CCA and with automated low flow anesthesia (LFA) (= maintenance FGF of 0.7 L min(-1)) with 3 different carrier gases.
Methods:
After IRB approval, 65 ASA PS I or II patients undergoing general surgery received desflurane in either O2, O2/air, or O2/N2O, with the Zeus to maintain the end-expired concentration (FA) at 6, 6, and 4% and the F1O2 at 1.0, 0.6, and 0.4, respectively. In addition, patients were assigned to either automated CCA (O2 n = 11; O2/air n = 11; O2/N2O n = 11) or automated LFA (selected FGF 0.7 L min(-1)) (O2 n = 12; O2/air n = 11; O2/N2O n = 9). Demographics and desflurane consumption at 2, 4, 6, 8, 10, 20, 30, 40 and 50 min were compared.
Results:
With the same carrier gas, desflurane consumption was lower with the CCA mode than with LFA mode after 4 min in the O2 groups, 6 min in the O2/air groups, and 30 min in the O2/N2O groups. Within each mode, desflurane consumption in the O2 and O2/air groups was identical at all times. Despite the use of a lower FA in the N2O groups, initial desflurane consumption was higher than in the O2 and O2/air groups, but it was lower later (> or = 15 min) only with LFA.
Discussion:
After 50 min, desflurane consumption with automated CCA is lower than with automated LFA. However, initial agent consumption is complex, and N2O in particular may increase initial desflurane consumption (though ultimately resulting in lower desflurane usage because of its MAC sparing effect) because initial FGF is increased to rapidly reach the target concentrations. Differences in desflurane consumption only become apparent after FGF has stabilized to the target FGF.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Related Concept Videos
Inhalational Anesthetics: Overview
Stages of General Anesthesia
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Parenteral Anesthetics: Overview
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...