Determination of ethylglucuronide in oral fluid by ultra-performance liquid chromatography- tandem mass spectrometry
S Hegstad1, L Johnsen, J Mørland
1Norwegian Institute of Public Health, Division of Forensic Toxicology and Drug Abuse, Nydalen, NO-0403 Oslo, Norway. solfrid.hegstad@stoalv.no
Abstract:
An ultra-performance liquid chromatography-tandem mass spectrometry method has been developed and validated for the determination of ethylglucuronide (EtG) in oral fluid. Sample clean-up was achieved by solid-phase extraction with a Hyper-SEP SAX column. Negative ionization was performed in the multiple reaction monitoring mode. Two transitions were monitored for the analyte and one for the internal standard EtG-d(5). The calibration range was 4.4-222 ng/mL. The recovery of the analyte ranged from 86 to 99%, and the between-assay precisions ranged from 5 to 9% RSD. The limit of quantification was found to be 4.4 ng/mL. The concentration of EtG in oral fluid collected 2-14 h after a moderate alcohol intake varied from 13.3 to 57.7 ng/mL.
More Related Videos
11:06GC-based Detection of Aldononitrile Acetate Derivatized Glucosamine and Muramic Acid for Microbial Residue Determination in Soil
Published on: May 19, 2012
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024