Benefit of higher closed-loop bandwidths in ocular adaptive optics
Abstract:
We present an ocular adaptive optics system with a wavefront sampling rate of 240 Hz and maximum recorded closed-loop bandwidth close to 25 Hz, but with typical performances around 10 Hz. The measured bandwidth depended on the specific system configuration and the particular subject tested. An analysis of the system performance as a function of achieved bandwidth showed consistently higher Strehl ratios for higher closed-loop bandwidths. This may be attributed to a combination of limitations on the available technology and the dynamics of ocular aberrations. We observed dynamic behaviour with a maximum frequency content around 30 Hz.