High risk infusions--accuracy compromised by changes in patient venous pressure
1Faculty of Engineering, Manukau Institute of Technology, Auckland, New Zealand. murphy16480@hotmail.com
Abstract:
Intensive care often requires the simultaneous administration of high potency short half-life intravenous drugs. Numerous reported instances of unwanted transient flow characteristics such as bolus flow, slow start-up at set rate and fluid reflux-where fluid is found to flow away from the infusion site-have indicated a need for better understanding of the system dynamic. This article introduces a simple mathematical model of an infusion pump system, highlighting system mechanical compliance as a major contributory factor for flow error. The model is verified as correctly predicting flow variation caused by pump height change and is offered as an aid in the development of more accurate infusion devices. The accurate delivery of low flow rates requires small volume rigid components. This study also suggests that the current design of syringe infusion devices makes them susceptible to flow error caused by change in patient venous pressure.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Venous Thrombosis IV: Nursing Management
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Venous Thrombosis III: Interprofessional Care
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies