Related Experiment Video
Updated: Jun 22, 2026

Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Radiation dose exposure of computed tomography coronary angiography: comparison of dual-source, 16-slice and 64-slice
1Department of Cardiology, Kerckhoff Heart Centre, Bad Nauheim, Germany. j.rixe@kerckhoff-klinik.de
Background:
Dual-source CT (DSCT) promises a significant reduction of radiation dose exposure for coronary CT angiography (CTA). Large studies on radiation dose estimates are rare.
Objective:
To compare radiation dose estimates of DSCT with 16- and 64-slice multidetector CT (MDCT) for non-invasive coronary angiography.
Patients And Design:
Retrospective data analysis was performed on 292 patients: 56 patients were examined with 16-slice MDCT, 38 patients with 64-slice MDCT and 202 patients using DSCT. The effective dose (ED) estimates were calculated for all patients from the dose-length product and the conversion factor k (0.017 mSv/mGy/cm), as recommended by current guidelines.
Results:
The mean (SD) ED for patients examined by 16-slice MDCT was 9.8 (1.8) mSv, for 64-slice MDCT 8.6 (2.8) mSv and for DSCT 11.4 (7.2) mSv. With a protocol of 100 kV tube voltage and 110 ms ECG pulsing window the mean (SD) ED was 3.8 (1.7) mSv for DSCT scanning. When DSCT with a tube voltage of 100 kV was used, a significant inverse correlation between heart rate and radiation dose exposure was found.
Conclusions:
When standard protocols for coronary CTA with 16-, 64-slice MDCT and DSCT scanners are used, the radiation dose is still high. However, using optimised and individually adjusted protocols low estimated radiation doses can be achieved.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT