Related Experiment Video
Updated: Jun 22, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Cementless acetabular socket revisions using Metasul metal-on-metal bearings
Kyung Soon Park1, Taek Rim Yoon, Eun Kyoo Song
1Center for Joint Disease, Chonnam National University Hwasun Hospital, Hwasun-Gun, Jeonnam, Korea.
Abstract:
Thirty-four hips were treated with cementless acetabular socket revisions using a metal-on-metal bearing. The causes of revision were aseptic loosening in 33 hips and septic loosening in 1 hip. Revisions were performed for acetabular sockets in 28 hips and for acetabular sockets and femoral stems in 6 hips. Mean follow-up duration was 6.2 years (range, 4.0-9.1 years), and mean Harris Hip Scores improved from 56 to 92 points. No hip required further revision for aseptic loosening. Focal femoral osteolysis newly developed in zone I in 1 hip, which was treated by curettage and bone grafting. The authors suggest that second-generation metal-on-metal bearings in cementless acetabular socket revisions can achieve good medium-term clinical and radiographic results.