Comparison of a commercial method for total protein with a candidate reference method
P D Camara1, C Wright, P Dextraze
1Roger Williams Hospital, Department of Pathology and Laboratory Medicine, Providence, RI 02908.
Annals of Clinical and Laboratory Science
|September 1, 1991
Summary
The Boehringer Mannheim Diagnostics Hitachi 717 Analyzer
Area of Science:
- Clinical Chemistry
- Analytical Biochemistry
Background:
- Standardization of total protein measurement is crucial for clinical diagnostics.
- The biuret method is a widely used assay for total protein determination.
- Evaluating automated analyzers against reference methods ensures accuracy and reliability.
Purpose of the Study:
- To standardize the biuret method for total protein measurement on the Boehringer Mannheim Diagnostics (BMD) Hitachi 717 Analyzer.
- To compare the BMD Hitachi 717 biuret method with a candidate reference method.
- To assess the accuracy, precision, and linearity of the BMD Hitachi 717 biuret method.
Main Methods:
- Comparison of the BMD Hitachi 717 biuret method with a candidate reference method.
- Analysis of 115 paired serum specimens.
- Statistical procedures for quantifying bias and linear regression analysis were employed.
Main Results:
- The Hitachi 717 biuret method demonstrated no significant bias compared to the reference method (z = -0.90).
- Acceptable correlation was observed between the two methods (r = 0.896).
- The method exhibited excellent precision (CV ≤ 2.14%) and linearity up to 15.0 g/dL.
Conclusions:
- The Boehringer Mannheim Diagnostics Hitachi 717 Analyzer biuret method is accurate and reliable for total protein measurement.
- The method is recommended for routine clinical use due to its convenience and accuracy.
- The standardization efforts confirm the suitability of the BMD Hitachi 717 for total protein analysis.