Background:
Cardiac CT angiography (CCTA) has become a frequently used diagnostic tool in clinical practice, but concern remains about the radiation exposure. Because of the second x-ray acquisition system, dual-source CT systems might allow for high-pitch CT data acquisition and thus for examination of the whole heart during a single heart beat, with the potential for radiation dose reduction.
Objective:
We assessed the feasibility of a high-pitch scan mode with a dual-source CT system.
Methods:
High-pitch modes were used in patients undergoing CCTA with a dual-source CT system. Diagnostic image quality for cardiac structures and coronary arteries was assessed. Radiation dose was estimated from the scanner-generated dose-length product (DLP).
Results:
CCTA was performed in 14 patients during a single heart beat applying a pitch value of 3.4. Mean heart rate during examination was 56.4+/-8.1 beats/min. Diagnostic image quality for the assessment of larger cardiac structures was obtained in all patients, whereas diagnostic image quality could be achieved in 82% of all coronary segments. With a mean DLP of 145+/-47 mGy x cm, the resulting estimated radiation dose was 2.0+/-0.7 mSv.
Conclusions:
This proof-of-concept study shows the ability of dual-source CT scanners to scan the whole heart during one single heart beat at low radiation dose.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...