Suppressing phase errors from vibration in phase-shifting interferometry

Leslie L Deck1

  • 1Zygo Corporation, Middlefield, Connecticut 06455, USA. LDeck@zygo.com

Applied Optics
|July 14, 2009
PubMed

Related Concept Videos

Time and frequency -Domain Interpretation of Phase-lead Control01:24

Time and frequency -Domain Interpretation of Phase-lead Control

Phase-lead controllers are commonly used in various control systems to enhance response speed and stability. Adjusting the brightness on a television screen offers a practical example of phase-lead control. When contrast is enhanced, a phase-lead controller is employed. Mathematically, phase-lead control is identified when the first parameter is smaller than the second.
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Phase Contrast and Differential Interference Contrast Microscopy01:26

Phase Contrast and Differential Interference Contrast Microscopy

Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Time and frequency -Domain Interpretation of Phase-lag Control01:21

Time and frequency -Domain Interpretation of Phase-lag Control

Phase-lag controllers are widely used in control systems to improve stability and reduce steady-state errors. A dimmer switch controlling the brightness of a light bulb serves as a practical example of phase-lag control, gradually adjusting the bulb's brightness. Mathematically, phase-lag control or low-pass filtering is represented when the factor 'a' is less than 1.
Phase-lag controllers do not place a pole at zero, but instead influence the steady-state error by amplifying any finite,...
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations01:08

IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations

Identical bonds within a polyatomic group can stretch symmetrically (in-phase) or asymmetrically (out-of-phase). Similar to hydrogen bonding, these vibrations also influence the shape of the IR peak. Generally, asymmetric stretching frequencies are higher than symmetric stretching frequencies. For example, primary amines exhibit two distinct IR peaks between 3300–3500 cm−1 corresponding to the symmetric and asymmetric N-H stretching, while secondary amines exhibit a single stretching vibration...
Propagation of Uncertainty from Systematic Error01:10

Propagation of Uncertainty from Systematic Error

The atomic mass of an element varies due to the relative ratio of its isotopes. A sample's relative proportion of oxygen isotopes influences its average atomic mass. For instance, if we were to measure the atomic mass of oxygen from a sample, the mass would be a weighted average of the isotopic masses of oxygen in that sample. Since a single sample is not likely to perfectly reflect the true atomic mass of oxygen for all the molecules of oxygen on Earth, the mass we obtain from this particular...
NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...