Related Experiment Video
Updated: Jun 21, 2026

A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Performance of a Chromolith RP-18e column for the screening of beta-blockers
Sandra Pous-Torres1, Maria-José Ruiz-Angel, José Ramón Torres-Lapasió
1Departament de Química Analítica, Universitat de València, Burjassot, Spain.
Abstract:
The chromatographic performance of a monolithic column (Chromolith RP-18e) was comprehensively examined in the isocratic separation of ten beta-blockers, using ACN-water mobile phases, and compared with the performance of three microparticulate RP columns manufactured with different types of silica: Spherisorb ODS-2, Kromasil C18 and XTerra MS C18. The comparison considered the analysis time, selectivity, peak shape (column efficiency and asymmetry) and resolution, and was extended to a wide range of mobile phase compositions. The Chromolith column showed good performance for the analysis of beta-blockers with regard to the packed columns. In terms of selectivity and analysis time, the greatest similarity was found between the Chromolith and XTerra columns. The addition of a silanol blocking agent (0.1% triethylamine) to both Chromolith and Spherisorb columns yielded, apparently, a similar blocking degree of the silanol groups (based on the similar peak shapes), and gave rise to similar selectivity.