Related Experiment Video
Updated: Jun 21, 2026

Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
A simple method to measure proton beam energy in a standard medical cyclotron
J W Burrage1, A H Asad, R A Fox
1Department of Medical Engineering and Physics, Royal Perth Hospital, Perth, Australia. John.Burrage@health.wa.gov.au
Abstract:
A simple and rapid technique to measure the proton beam energy in the external beam line of a medical cyclotron has been examined. A stack of 0.1 mm thick high purity copper (Cu) foils was bombarded and the relative activity of 65Zn produced in each foil was compared to a computational model that predicted activity, based on proton stopping power, reaction cross-sectional data, and beam energy. In the model, the beam energy was altered iteratively until the best match between computed and measured relative activities of the stack of disks was obtained. The main advantage of this method is that it does not require the comparison of the activities of different isotopes of zinc arising from (p, xn) reactions in the Cu, which would require the gamma photon detector being calibrated for different energy responses. Using this technique the proton beam energy of a nominally 18 MeV standard isochronous medical cyclotron was measured as 17.49 +/- 0.04 (SD) MeV, with a precision of 0.2% CV.
Related Concept Videos
Mass Analyzers: Common Types
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei in a...
Nuclear Transmutation
Biological Effects of Radiation