Related Experiment Video
Updated: Jun 21, 2026

Rapid Formation and Testing of Self-expanding NiTi Frames with a Small Form Factor Suitable for Minimally Invasive Implants
Published on: March 7, 2025
Are more nickel ions released from NiTi wires after sterilisation?
Maryam Poosti1, Hossein Pezeshki Rad, Kamyar Kianoush
1Orthodontic Department, School of Dentistry and Dental Research Center, Mashhad University of Medical Sciences, Iran. poustim@mums.ac.ir
Background:
Increased concentrations of nickel ions may be released into saliva from recycled NiTi archwires.
Aim:
To compare the nickel ion concentrations released from recycled NiTi wires after sterilisation by either dry heat or steam autoclave.
Methods:
Eighty preformed NiTi wires were assigned to four equal groups. In Groups 1, 2 and 3 the archwires were used intra-orally for 4 weeks. The Group 4 archwires were not used. Group 1 archwires were sterilised by dry heat, the Group 2 archwires were sterilised by steam autoclave, the Group 3 archwires were not sterilised and the Group 4 archwires were as-received. A 2 cm length, cut from each archwire, was immersed in artificial saliva for 4 weeks and the nickel ion concentrations in the artificial saliva measured with an atomic absorption spectrophotometer.
Results:
There were no significant differences in the nickel ion concentrations released into the artificial saliva by each group of archwires (p = 0.467).
Conclusion:
Sterilisation of used NiTi wires by either dry heat and steam autoclave does not affect the concentrations of nickel ions released into artificial saliva.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
2° Amines to N-Nitrosamines: Reaction with NaNO2
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...
Electrodeposition
Electrodeposition can...
Formation of Complex Ions