Background:
We undertook a pilot investigation to evaluate the feasibility of a novel technique using intracardiac echocardiography (ICE) for intraoperative assessment of cardiac resynchronization therapy (CRT).
Methods:
We evaluated ICE intraoperative imaging of left ventricular (LV) function and aortic valvular flow as well as safety of implementation. ICE was used to guide CRT system lead placement, assess impact of pacing modes, and optimization of device programming.
Results:
Twenty-three patients underwent ICE imaging. ICE showed global hypokinesis in six patients, regional wall motion abnormality only in 10 patients, and both in seven patients. Optimized CRT modes included mean atrioventricular (AV) interval of 170 ms and interventricular timing using simultaneous right ventricular (RV)-LV pacing (five patients), LV pacing only (one patient), and sequential LV to RV stimulation (15 patients) or RV to LV stimulation (two patients). ICE-guided CRT acutely improved mean left ventricular ejection fraction (LVEF) from 24 +/- 9% to 41 +/- 1% (P < 0.00001). During follow-up of 3-24 (mean 11) months, New York Heart Association class improved in all patients from a mean of 3.2 +/- 0.4 at implant to 1.6 +/- 0.7 (P < 0.0001), with improvement of LVEF from 19 +/- 7% to 34 +/- 12% (P = 0.0001). Actuarial survival was 83% at 12 months.
Conclusions:
(1) ICE imaging is reliable and safe for continuous intraoperative imaging of LV wall motion, and assesses baseline status and impact of CRT interventions.(2) Intraoperative ICE-guided CRT optimization resulted in an increase in LVEF acutely and consistent improvement in heart failure.(3) Sequential biventricular pacing and longer AV interval programming were more often used in ICE-guided CRT.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:43Advanced Cardiac Rhythm Management by Applying Optogenetic Multi-Site Photostimulation in Murine Hearts
Published on: August 26, 2021
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Heart Failure VI: Adjunct Therapies
Cardiomyopathy II: Dilated Cardiomyopathy
Dysrhythmias VI: Management of Dysrhythmias
Cardiomyopathy IV: Restrictive Cardiomyopathy
Cardiopulmonary Resuscitation III: AED Use