Related Experiment Video
Updated: Jun 21, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
ICD proarrhythmia as a consequence of an interaction with an algorithm to prevent atrial arrhythmias
Juan Martínez-Sánchez1, Arcadi García-Alberola, Juan José Sánchez-Muñoz
1Department of Cardiology, University Hospital Virgen de la Arrixaca, Murcia, Spain. juanmsmur@secardiología.es
Abstract:
An episode of device-related proarrhythmia is reported. Ventricular tachycardia was induced by ventricular pacing in a dual-chamber implantable cardioverter-defibrillator due to an inappropriate interaction of programmed device settings with the atrial preference pacing, an automatic algorithm designed to prevent atrial tachyarrhythmias.