On the resolution capabilities and limits of fluorescence lifetime correlation spectroscopy (FLCS) measurements
Steffen Rüttinger1, Peter Kapusta, Matthias Patting
1Physikalisch-Technische Bundesanstalt, Abbestr. 2-12, 10587, Berlin, Germany. ruettinger@gmail.com
Abstract:
Quantitative tests were performed in order to explore the practical limits of FLCS. We demonstrate that: a) FLCS yields precise and correct concentration values from as low as picomolar to micromolar concentrations; b) it is possible to separate four signal components in a single detector setup; c) diffusion times differing only 25% from each other can be resolved by separating a two component mixture based on the different fluorescence lifetimes of both components; d) most of the inherent technical limitations of conventional FCS are easily overcome by FLCS employing a single detector channel confocal detection scheme.
More Related Videos
07:15Measuring Interactions between Fluorescent Probes and Lignin in Plant Sections by sFLIM Based on Native Autofluorescence
Published on: January 2, 2020
14:12Dual-Color Fluorescence Cross-Correlation Spectroscopy to Study Protein-Protein Interaction and Protein Dynamics in Live Cells
Published on: December 11, 2021
Related Concept Videos
Fluorescence and Phosphorescence: Instrumentation
Confocal Fluorescence Microscopy
Super-resolution Fluorescence Microscopy
Protein Dynamics in Living Cells
Fluorescent recovery after photobleaching (FRAP) is a fluorescent-protein-based detection technique used to quantify protein movement rates within the cell. This method exposes a small portion of the cell to an intense laser beam. The laser beam causes permanent photobleaching of the fluorophore-tagged proteins in the exposed region. As the bleached...