Related Experiment Video
Updated: Jun 20, 2026

Implementation of a Real-Time Psychosis Risk Detection and Alerting System Based on Electronic Health Records using CogStack
Published on: May 15, 2020
Alarm limit settings for early warning systems to identify at-risk patients
Lawrence P A Burgess1, Tracy Heather Herdman, Benjamin W Berg
1Hoana Medical, Inc. Honolulu, Hawaii, USA. lburgess@hoana.com
Aim:
This paper is a report of a study conducted to provide objective data to assist with setting alarm limits for early warning systems.
Background:
Early warning systems are used to provide timely detection of patient deterioration outside of critical care areas, but with little data from the general ward population to guide alarm limit settings. Monitoring systems used in critical care areas are known for excellent sensitivity in detecting signs of deterioration, but give high false positive alarm rates, which are managed with nurses caring for two or fewer patients. On general wards, nurses caring for four or more patients will be unable to manage a high number of false alarms. Physiological data from a general ward population would help to guide alarm limit settings.
Methods:
A dataset of continuous heart rate and respiratory rate data from a general ward population, previously collected from July 2003-January 2006, was analyzed for adult patients with no severe adverse events. Dataset modeling was constructed to analyze alarm frequency at varying heart rate and respiratory rate alarm limits.
Results:
A total of 317 patients satisfied the inclusion criteria, with 780.71 days of total monitoring. Sample alarm settings appeared to optimize false positive alarm rates for the following settings: heart rate high 130-135, low 40-45; respiratory rate high 30-35, low 7-8. Rates for each selected limit can be added to calculate the total alarm frequency, which can be used to judge the impact on nurse workflow.
Conclusion:
Alarm frequency data will assist with evidence-based configuration of alarm limits for early warning systems.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Holter Monitor: 24-Hour Monitoring