Accuracy and precision of linear measurements in cone beam computed tomography Accuitomo tomograms obtained with
H Lund1, K Gröndahl, H-G Gröndahl
1Department of Oral and Maxillofacial Radiology, Sahlgrenska Academy, University of Gothenburg, Sweden. henrik.lund@vgregion.se
Objectives:
To assess the accuracy and precision of linear measurements, the influence of different object positions in cone beam CT (CBCT) images and the effect of different systems for reformatting and viewing the tomograms on measurement values.
Methods:
An object consisting of Plexiglas plates with metal spheres was radiographically examined in three positions using the CBCT unit Accuitomo. Tomograms were reformatted using the i-Dixel software and Sectra picture archiving and communication system (PACS) multiplanar reformatting (MPR). Two observers measured 20 linear distances in axial, frontal and sagittal planes. Physical measurements of actual distances by means of a digital caliper served as the gold standard.
Results:
For measurements performed on the Accuitomo and Sectra PACS workstations, the mean differences between repeated measurements were below 0.13 mm (standard deviation (SD) 0.13 mm) and 0.21 mm (SD 0.18 mm), respectively. Mean differences between the two observers were below 0.11 mm (SD 0.07 mm) and 0.15 mm (SD 0.09 mm). The mean difference between the gold standard and the basic position when using the Accuitomo workstation, was -0.08 mm (SD 0.21 mm) and for a deviated/rotated position -0.08 mm (SD 0.23 mm) and -0.09 mm (SD 0.20 mm). None was statistically significant (P > 0.05). For measurement performed on MPR tomograms using the Sectra PACS workstation, mean differences were -0.09 mm (SD 0.17 mm) for basic position, and -0.08 mm (SD 0.19 mm) and -0.13 mm (SD 0.15 mm) for deviated and rotated positions, respectively. The mean difference between gold standard and basic/rotated positions was statistically significant, with P = 0.032 and P = 0.002.
Conclusions:
Our study confirms previous findings indicating that measurements in 3DX Accuitomo FPD (flat panel detector) tomograms show a high level of agreement with actual distances. Our findings suggest the possibility of minor distortion in tomograms reformatted by means of Sectra PACS MPR.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Imaging Studies III: Computed Tomography
Accuracy and Precision
Accuracy and Precision