Analysis of ureteric stent kinking forces: the role of curvature in stent failure
Matthew S Christman1, James O L'Esperance, Chong H Choe
1Department of Urology, Naval Medical Center San Diego, San Diego, CA, USA.
Objective:
To further characterize the kinking forces and degree of curvature that result in failure of various stents, as malignant obstruction of the ureter causes failure of about half of ureteric stents.
Materials And Methods:
Several stents (Silhouette 4.6 F and 6 F, Applied Medical, Rancho Santa Margarita, CA; Sof-Curl Tecoflex 6 F, Gyrus ACMI, Southborough, MA; Polaris Ultra 6 and 7 F, and Percuflex 6, 7 and 8 F stents, Boston Scientific, Natick, MA, USA) were tested. The amount of force necessary to result in kinking of the stent was measured, and the degree of curvature at failure was calculated for each stent.
Results:
The Silhouette 4.6 and 6 F stents were the most resistant to failure by kinking and curvature. In general, smaller stents allowed more curvature before failing than their larger counterparts.
Conclusions:
The greater allowable curvature and resistance to kinking achieved by the Silhouette ureteric stents might result in fewer stent failures in cases of malignant obstruction or other retroperitoneal processes.
Insights
Silhouette ureteric stents demonstrate superior resistance to kinking and curvature, potentially reducing failures in malignant ureteral obstructions. Smaller stents generally allow more bending before failure.
Area of Science:
- Urology
- Biomedical Engineering
- Medical Device Design
Background:
- Ureteral stent failure is common, with malignant obstruction causing approximately 50% of failures.
- Kinking and excessive curvature are primary modes of ureteric stent failure.
Purpose of the Study:
- To quantify the kinking forces and degree of curvature leading to failure in various ureteral stents.
- To identify stent designs with improved resistance to mechanical failure.
Main Methods:
- Mechanical testing of multiple ureteral stent models (Silhouette, Sof-Curl Tecoflex, Polaris Ultra, Percuflex) with varying French sizes.
- Measurement of the force required to induce stent kinking.
- Calculation of the degree of curvature at the point of stent failure.
Main Results:
- Silhouette 4.6 F and 6 F stents exhibited the highest resistance to kinking and curvature-induced failure.
- Smaller diameter stents generally tolerated greater degrees of curvature before failure compared to larger stents.
Conclusions:
- Silhouette ureteric stents offer enhanced resistance to kinking and greater allowable curvature.
- These properties may lead to reduced stent failure rates in patients with malignant ureteral obstruction and other retroperitoneal conditions.
Related Concept Videos
Bending of Curved Members - Strain Analysis
The important part of bending analysis for such a member is the...
Residual Stresses in Bending
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Stress-Strain Diagram - Ductile Materials
Deformations in a Symmetric Member in Bending
When the member is segmented into tiny cubic elements, it is observed that the primary stress...
Plastic Deformations