Related Experiment Video
Updated: Jun 20, 2026

Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Effect of transcatheter pulmonary valve implantation on short-term right ventricular function as determined by
Nasser Moiduddin1, Kentaro Asoh, Cameron Slorach
1Division of Paediatric Cardiology, The Labatt Family Heart Center, The Hospital for Sick Children, Toronto, Ontario, Canada.
Insights
Transcatheter pulmonary valve implantation (PVI) improved right ventricular (RV) function in children with congenital heart disease. This emerging therapy reduced RV pressure and improved RV strain and septal function shortly after the procedure.
Area of Science:
- Cardiology
- Pediatric Cardiology
- Echocardiography
Background:
- Transcatheter pulmonary valve implantation (PVI) is an innovative treatment for right ventricular (RV) outflow dysfunction in pediatric congenital heart disease.
- Assessing the immediate impact of PVI on RV and left ventricular (LV) function is crucial for understanding its therapeutic benefits.
Purpose of the Study:
- To evaluate the short-term effects of PVI on RV and LV function in children post-congenital heart surgery.
- To investigate changes in myocardial mechanics using 2D speckle tracking echocardiography and tissue Doppler imaging.
Main Methods:
- Utilized 2D speckle tracking echocardiography and tissue Doppler imaging to assess RV and LV function before and after PVI.
- Measured strain and strain rate of the LV, RV free wall, and interventricular septum.
- Recorded catheter right heart hemodynamics and pulmonary regurgitation.
Main Results:
- PVI significantly reduced the RV to pulmonary artery pressure gradient and RV end-diastolic volume.
- Septal systolic velocities, strain, and strain rate, along with RV free wall strain, showed significant improvement post-PVI.
- Left ventricular function, assessed by tissue velocities, strain, and strain rate, remained unchanged.
Conclusions:
- Transcatheter pulmonary valve implantation effectively unloads the right ventricle.
- PVI acutely enhances right ventricular and septal myocardial function in pediatric patients with congenital heart disease.
Abstract:
Transcatheter pulmonary valve implantation (PVI) is an emerging therapy for right ventricular (RV) outflow dysfunction in congenital heart disease. We investigated, for the first time in children after surgery for congenital heart disease, the short-term effects of PVI on RV and left ventricular (LV) function using 2-dimensional speckle tracking echocardiography and tissue Doppler imaging. We hypothesized that the short-term RV and LV function would improve. Two-dimensional speckle tracking echocardiograms and pulsed tissue Doppler images were obtained before and 1 to 2 days after PVI (18-mm Melody valve). The catheter right heart hemodynamics were recorded. The strain and strain rate of the basal lateral left ventricle, lateral right ventricle, and interventricular septum were measured by 2-dimensional speckle tracking echo, and the pre- and postprocedure values were compared. Of the 16 eligible patients (age 16 +/- 2 years), the scans of 10 had correct image format and adequate quality. PVI was performed for volume (n = 4) or combined pressure-volume (n = 6) loading. After PVI, the RV to pulmonary artery pressure gradient (33 +/- 22 to 12 +/- 4 mm Hg, p = 0.02), pulmonary regurgitation, and RV end-diastolic volume (3.2 +/- 0.8 to 2.8 +/- 0.6 cm, p = 0.02) decreased, and the septal systolic velocities (3.5 +/- 1.1 to 4.7 +/- 1.1 cm/s, p = 0.04), strain (-7.6 +/- 9.3% to -15.6% +/- 6.7%, p = 0.01) and strain rate (-0.3 +/- 1.1 to -1.1 +/- 0.5 1/s, p = 0.04) and RV free wall strain increased (-17.4 +/- 8.6% to -23.4% +/- 6.2%, p = 0.03). The LV tissue velocities, strain, and strain rate were unchanged. In conclusion, PVI leads to RV unloading and acutely improves RV and septal function.
More Related Videos
10:33Assessment of Right Ventricular Structure and Function in Mouse Model of Pulmonary Artery Constriction by Transthoracic Echocardiography
Published on: February 3, 2014
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022