Rapid determination of telmisartan in human plasma by HPLC using a monolithic column with fluorescence detection and
He Zhang1, Yunyun Jiang, Jun Wen
1Department of Pharmaceutical Analysis, School of Pharmacy, Second Military Medical University, No. 325 Guohe Road, Shanghai 200433, PR China.
Abstract:
A rapid HPLC method using a monolithic column with fluorescence detection has been developed for determination of telmisartan in human plasma. Sample preparation was done by protein precipitation with acetonitrile and naproxen was used as IS. The compounds were detected by fluorescence detection, using an excitation wavelength of 300 nm and emission wavelength of 385 nm. Calibration curves of telmisartan were linear in the range of 1-200 ng/mL. The assay was high throughput, sensitive and precise, and it was successfully applied to a bioequivalence study of two formulations of telmisartan.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Elution Process
Bioequivalence of Drugs: Drugs with Multiple Indications