Proposal of a method for the evaluation of inaccuracy of home sphygmomanometers
1Department of Nephrology, Ondokuz Mayis University School of Medicine, Samsun, Turkey. tekinakpolat@yahoo.com
Objective:
There is no formal protocol for evaluating the individual accuracy of home sphygmomanometers. The aims of this study were to propose a method for achieving accuracy in automated home sphygmomanometers and to test the applicability of the defined method. The purposes of this method were to avoid major inaccuracies and to estimate the optimal circumstance for individual accuracy.
Methods:
The method has three stages and sequential measurement of blood pressure is used. The tested devices were categorized into four groups: accurate, acceptable, inaccurate and very inaccurate (major inaccuracy). The defined method takes approximately 10 min (excluding relaxation time) and was tested on three different occasions.
Results:
The application of the method has shown that inaccuracy is a common problem among non-tested devices, that validated devices are superior to those that are non-validated or whose validation status is unknown, that major inaccuracy is common, especially in non-tested devices and that validation does not guarantee individual accuracy.
Conclusion:
A protocol addressing the accuracy of a particular sphygmomanometer in an individual patient is required, and a practical method has been suggested to achieve this. This method can be modified, but the main idea and approach should be preserved unless a better method is proposed. The purchase of validated devices and evaluation of accuracy for the purchased device in an individual patient will improve the monitoring of self-measurement of blood pressure at home. This study addresses device inaccuracy, but errors related to the patient, observer or blood pressure measurement technique should not be underestimated, and strict adherence to the manufacturer's instructions is essential.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.