Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Detection of carbon dioxide thresholds using low-flow sidestream capnography in ventilated preterm infants
Emmanuel Lopez1, Sophie Grabar, Alexandre Barbier
1Service de Médecine néonatale, Groupe Hospitalier Cochin-Saint Vincent de Paul, AP-HP, 123 Bd du Port-Royal, 75014 Paris, France. emmanuel.lopez@cch.aphp.fr
Background:
Monitoring CO2 levels in preterm infants receiving mechanical ventilation is designed to avoid the harmful consequences of hypocapnia or hypercapnia. Capnography is of questionable accuracy for monitoring PCO2 in preterm infants.
Objectives:
To determine the accuracy of sidestream capnography in ventilated preterm infants by comparing end-tidal carbon dioxide (EtCO2) values to mixed venous carbon dioxide pressure (PvCO2) and to transcutaneous carbon dioxide pressure (TcPCO2).
Methods:
Simultaneous recordings of EtCO2, TcPCO2 and PvCO2 in 37 ventilated preterm infants. The PvCO2-EtCO2 gradient was calculated. The Bland-Altman technique and the intra-class correlation coefficient (ICC) were used to assess agreement between methods. The area under the curve (AUC) was calculated.
Results:
Ninety-nine EtCO2/PvCO2 pairs were studied from 37 preterm infants with a mean gestational age of 27.7 +/- 1.9 weeks and a mean birth weight of 1,003 +/- 331 g. The mean PvCO2-EtCO2 gradient was 11.2 +/- 8.0 mmHg, and the ICC was 0.28. The mean PvCO2-TcPCO2 gradient was 0 +/- 7.8 mmHg, and the ICC was 0.78. AUCs for EtCO2 and TcPCO2 were similar in detecting high or low PvCO2.
Conclusion:
Despite an insufficient correlation between EtCO2 and PvCO2, capnography was able to detect low and high CO2 warning levels with a similar efficacy to that of TcPCO2, and may therefore be of clinical interest.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Acute Respiratory Failure-III
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)