Radiation dose for body CT protocols: variability of scanners at one institution
Tracy A Jaffe1, Terry T Yoshizumi, Greta Toncheva
1Department of Radiology, Duke University Medical Center, Erwin Rd., Box 3808, Durham, NC 27710, USA. jaffe002@mc.duke.edu
Objective:
The objective of our study was to determine, using an anthropomorphic phantom, whether patients are subject to variable radiation doses based on scanner assignment for common body CT studies.
Materials And Methods:
Twenty metal oxide semiconductor field effect transistor dosimeters were placed in a medium-sized anthropomorphic phantom of a man. Pulmonary embolism and chest, abdomen, and pelvis protocols were used to scan the phantom three times with GE Healthcare scanners in four configurations and one 64-MDCT Siemens Healthcare scanner. Organ doses were averaged, and effective doses were calculated with weighting factors.
Results:
The mean effective doses for the pulmonary embolism protocol ranged from 9.9 to 18.5 mSv and for the chest, abdomen, and pelvis protocol from 6.7 to 18.5 mSv. For the pulmonary embolism protocol, the mean effective dose from the Siemens Healthcare 64-MDCT scanner was significantly lower than that from the 16- and 64-MDCT GE Healthcare scanners (p < 0.001). The mean effective dose from the GE 4-MDCT scanner was significantly lower than that for the GE 16-MDCT scanner (p < 0.001) but not the GE 64-MDCT scanner (p = 0.02). For the chest, abdomen, and pelvis protocol, all mean effective doses from the GE scanners were significantly different from one another (p < 0.001), the lowest mean effective dose being found with use of a single-detector CT scanner and the highest with a 4-MDCT scanner. For the chest, abdomen, and pelvis protocols, the difference between the mean effective doses from the GE Healthcare and Siemens Healthcare 64-MDCT scanners was not statistically significant (p = 0.89).
Conclusion:
According to phantom data, patients are subject to different radiation exposures for similar body CT protocols depending on scanner assignment. In general, doses are lowest with use of 64-MDCT scanners.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT