Correlation between screening for obstructive sleep apnea using a portable device versus polysomnography testing in a
Marc R Watkins1, James B Talmage, Matthew S Thiese
1Division of Occupational and Environmental Medicine, Meharry Medical College, Nashville, TN, USA. marcwatkins@comcast.net
Objective:
To compare the accuracy of portable monitoring for obstructive sleep apnea (OSA) with polysomnography (PSG) in commercial drivers.
Methods:
A consecutive case series of drivers were screened for OSA using recent validated consensus conference criteria at a single occupational medicine clinic; those screening positive were tested with a portable OSA screening device while awaiting PSG.
Results:
Three hundred forty-six drivers were screened and 34 completed PSG and portable monitoring. Comparing the portable device to PSG at a definition of OSA as an apnea hypopnea index (AHI) >or=15, positive predictive value is 0.64 with a 0.87 negative predictive value. Positive likelihood ratio is 4.20. Linear regression model found statistically significant relationship (P = 0.0004) between the portable device and PSG AHI values.
Conclusion:
Although not perfect, portable monitoring may be useful in further stratifying risk in drivers who have a high probability of having OSA.