Related Experiment Video
Updated: Jun 19, 2026

Morphological and Functional Assessment of the Right Ventricle Using 3D Echocardiography
Published on: October 28, 2020
A study of functional anatomy of aortic-mitral valve coupling using 3D matrix transesophageal echocardiography
Federico Veronesi1, Cristiana Corsi, Lissa Sugeng
1Department of Electronics, Computer Science and Systems, University of Bologna, Italy.
Background:
Mitral and aortic valves are known to be coupled via fibrous tissue connecting the two annuli. Previous studies evaluating this coupling have been limited to experimental animals using invasive techniques. The new matrix array transesophageal transducer provides high-resolution real-time 3D images of both valves simultaneously. We sought to develop and test a technique for quantitative assessment of mitral and aortic valve dynamics and coupling.
Methods And Results:
Matrix array transesophageal (Philips iE33) imaging was performed in 24 patients with normal valves who underwent clinically indicated transesophageal echocardiography. Custom software was used to detect and track the mitral and aortic annuli in 3D space throughout the cardiac cycle, allowing automated measurement of changes in mitral and aortic valve morphology. Mitral annulus surface area and aortic annulus projected area changed reciprocally over time. Mitral annulus surface area was 8.0+/-2.1 cm(2) at end-diastole and decreased to 7.7+/-2.1 cm(2) in systole, reaching its maximum (10.0+/-2.2 cm(2)) at mitral valve opening. Aortic annulus projected area was 4.1+/-1.2 cm(2) at end-diastole, then increased during isovolumic contraction reaching its maximum (4.8+/-1.3 cm(2)) in the first third of systole and its minimum (3.6+/-1.0 cm(2)) during isovolumic relaxation. The angle between the mitral and aortic annuli was maximum (136+/-13 degrees ) at end-diastole and decreased to its minimum value (129+/-11 degrees ) during systole.
Conclusions:
This is the first study to report quantitative 3D assessment of the mitral and aortic valve dynamics from matrix array transesophageal images and describe the mitral-aortic coupling in a beating human heart. This ability may have impact on patient evaluation for valvular surgical interventions and prosthesis design.
More Related Videos
Related Concept Videos
Mitral Stenosis II: Clinical features and Diagnostic Tests
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Mitral Regurgitation I: Introduction
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...