Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
High-pitch electrocardiogram-triggered computed tomography of the chest: initial results
Michael Lell1, Fabian Hinkmann, Katharina Anders
1Department of Radiology, University of Erlangen, Erlangen, Germany. michael.lell@uk-erlangen.de
Insights
A new high-pitch scan mode for dual-source computed tomography (CT) enables motion-artifact-free imaging of thoracic vessels and coronary arteries. This low-dose technique offers diagnostic image quality, particularly beneficial for patients with stable heart rates.
Area of Science:
- Radiology
- Cardiovascular Imaging
- Medical Physics
Background:
- Chest pain is a common emergency department presentation with diverse causes, some life-threatening.
- Electrocardiogram (ECG)-gated computed tomography angiography (CTA) of the thorax is a proposed cost-effective imaging method.
- Dual-source CT (DSCT) systems offer advanced capabilities for cardiac imaging.
Purpose of the Study:
- To introduce and evaluate a novel high-pitch scan mode for low-dose ECG-triggered CT angiography.
- To assess the feasibility and image quality of this technique in patients presenting with chest pain.
- To determine the radiation exposure associated with the new scanning protocol.
Main Methods:
- Twenty-four patients underwent thoracic CT angiography using a second-generation DSCT system with a high-pitch (3.2) scan mode.
- Data acquisition was ECG-triggered (50-60% of RR interval) with a temporal resolution of 75 ms, completing the scan in under 1 second.
- A triple-phase contrast injection protocol was employed, and image quality was evaluated using a 4-point scale.
Main Results:
- Motion-artifact-free visualization of the aorta and pulmonary vessels was achieved in all patients.
- Coronary artery image quality was diagnostic, with 70% of segments scoring grade 1 (no artifacts).
- Mean effective radiation doses were low, averaging 1.6 mSv at 100 kV and 3.2 mSv at 120 kV.
Conclusions:
- The high-pitch scan mode facilitates motion-artifact-free and accurate visualization of thoracic vessels.
- Diagnostic image quality of coronary arteries is achievable, especially in patients with low and stable heart rates.
- This technique offers a low-radiation-dose option for thoracic CT angiography in emergency settings.
Objectives:
Chest pain is one of the most frequent symptoms in the emergency department. A variety of different diseases, some of them acutely life threatening, can be the underlying cause. Electrocardiogram (ECG)-gated computed tomography angiography of the thorax has been proposed as a cost and time effective imaging technique for these patients. We describe a new high-pitch scan mode, which has been developed specifically for low-dose ECG-triggered computed tomography angiography using dual source computed tomography (CT).
Material And Methods:
Twenty-four patients were examined with this technique on a second generation dual source CT system. The scan mode uses a pitch of 3.2 to acquire a spiral CT data set of the complete thorax in less than 1 second with a temporal resolution of 75 ms (scan parameters: 128 x 0.6 mm collimation, 0.28 seconds gantry rotation time, 370 mAs at 100 kV [15 patients] and 320 mAs at 120 kV [9 patients], reconstructed slice thickness 0.6 mm, increment 0.4 mm). Data acquisition was prospectively triggered at 50% to 60% of the RR interval to cover the range over the heart in diastole. A triple phase contrast injection protocol (total volume: 80 mL) was used to optimize enhancement of the pulmonary and systemic arterial vessels. Image quality was evaluated using a 4-point scale (1 = absence of motion artifacts; 2 = slight motion artifacts, fully evaluable; 3 = motion artifacts, but evaluable; 4 = unevaluable) on a per-segment basis.
Results:
The patients had an average heart rate of 68 +/- 15 bpm (range: 43-111 bpm) during data acquisition. Motion artifact free visualization of the aorta and pulmonary vessels was possible in each case, of 344 coronary artery segments, 242 (70%) had an image quality score of 1, 60 segments (17%) a score of 2, 28 segments (8%) a score of 3, and 14 segments (4%) were rated as "unevaluable." In 17 patients (10 patients with a heart rate < or =60 bpm) all segments were evaluable. The average dose length product was 113 +/- 11 mGy x cm per scan (mean effective dose 1.6 +/- 0.2 mSv) at 100 kV and 229 +/- 31 mGy x cm per scan (mean effective dose 3.2 +/- 0.4 mSv) at 120 kV.
Conclusion:
Our initial results indicate that this high-pitch scan mode allows motion artifact free and accurate visualization of the thoracic vessels, and diagnostic image quality of the coronary arteries in patients with low and stable heart rates at a very low radiation exposure.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Acute Coronary Syndrome III: Diagnostic Studies