A procedure for evaluation of non-invasive blood pressure simulators
Gregor Gersak1, Ales Zemva, Janko Drnovsek
1Faculty of Electrical Engineering, University of Ljubljana, Ljubljana, Slovenia. gregor.gersak@fe.uni-lj.si
Abstract:
Non-invasive blood pressure (NIBP) simulators are used in clinical environment for quick checks of blood pressure monitors as a part of technical maintenance and health-care quality assurance system. They are also included in various tests within the procedures for testing NIBP monitors. In practice simulators are often subject to mechanical and electromagnetic shocks which could effect their measuring function. Our objective was to design a procedure for testing the reliability and quality of simulators in order to ensure reliable testing of NIBP monitors. Procedure for evaluation of NIBP simulators, consisting of a static and dynamic test, is proposed. Static test consisted of procedures derived from common electro-mechanical manometer calibration, while dynamic test included testing of repeatability of simulator's output. A commercial simulator was tested. Among others, the results indicated that evaluations of NIBP simulators should be performed regularly with a suitable time interval in order to track the metrological quality of the simulator in time. Acceptance criteria for a reliable simulator in both static and dynamic sense are proposed.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure