Related Experiment Video
Updated: Jun 19, 2026

Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
Published on: June 20, 2015
Test-retest variability of intraocular pressure and ocular pulse amplitude for dynamic contour tonometry: a
P Fogagnolo1, M Figus, P Frezzotti
1GB Bietti Foundation-IRCCS, Rome, Italy. fogagnolopaolo@googlemail.com
Aims:
To assess the test-retest variability of intraocular pressure (IOP) and ocular pulse amplitude (OPA) measurements utilising dynamic contour tonometry (DCT) and to evaluate possible influential factors.
Methods:
The study included 350 consecutive subjects (175 glaucoma, 175 control; one eye per subject) from seven European centres. IOP was measured once with a Goldmann applanation tonometer (GAT) and twice by DCT (DCT1, DCT2) in a randomised sequence. OPA was also recorded for both DCT measurements. Differences (DCT1-DCT2; OPA1-OPA2; GAT-DCT1; GAT-DCT2) were assessed using the t test. The intraclass coefficient of correlation (ICC) and coefficient of variation (CoV) for DCT and OPA were calculated.
Results:
DCT1 was 0.6+/-1.6 mm Hg higher than DCT2 (p<0.001); OPA1 was 0.1+/-0.7 mm Hg higher than OPA2 (p=0.02). Results were not influenced by randomisation test order. In both glaucoma and normal subjects, DCT and OPA showed ICC>0.90 and >0.76, and CoV=4.8-5.0% and 10.3-10.5%, respectively. DCT1 and 2 were 2.4+/-2.6 and 1.8+/-2.6 mm Hg higher respectively than GAT (p<0.001).
Discussion:
DCT test-retest variability was almost perfect for IOP and good for OPA. Tonometry measurements with DCT tended to be overestimated compared with GAT.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
