Oxygen delivery using neonatal self-inflating resuscitation bags without a reservoir
1Neonatal Service, University Hospital Sant Joan de Déu, Barcelona, Spain. mthio@hsjdbcn.org
Background:
Guidelines recommend avoidance of excessive oxygen administration during neonatal resuscitation. Blenders are used in some but not all hospitals. It has been suggested that self-inflating bags without a reservoir deliver around 40% oxygen and could be used to provide an inexpensive and effective technique of avoiding oxygen toxicity.
Objective:
To explore how much oxygen is delivered when using two different brands of neonatal self-inflating resuscitation bags without a reservoir.
Methods:
In a benchtop setting, the smallest non-disposable self-inflating bags from the Laerdal and Ambu ranges were tested. Oxygen concentration delivered by these devices under a variety of conditions was measured. 108 combinations of oxygen flow rates (10; 5 to 1 litre/min), ventilation rates (30, 60, 100 inflations/min) and peak inspiratory pressure ranges (20 to 25 cm H(2)O, 35 to 40 cm H(2)O or pop-off valve range, 55 to 60 cm H(2)O) were tested.
Results:
Delivered oxygen concentration varied depending on three parameters: gas flow rate, ventilatory rate and pressure. At a pressure of 20 to 25 cm H(2)O, mean oxygen concentration delivered by both bags exceeded 70% at any gas flow rate except for 1 litre/min (where delivered oxygen concentration was 60% to 70%). When the pop-off valve was opened at 35 to 40 cm H(2)0, oxygen concentrations fell to 30% to 45% at gas flow rates < or =2 litres/min. The Ambu bag delivered a lower oxygen concentration than the Laerdal bag but this difference was not clinically important.
Conclusion:
When using the Laerdal and Ambu infant resuscitation self-inflating bags without a reservoir, delivered oxygen concentration is >70% for currently recommended flow and pressure settings.
More Related Videos
05:45Delivery of In Vivo Acute Intermittent Hypoxia in Neonatal Rodents to Prime Subventricular Zone-derived Neural Progenitor Cell Cultures
Published on: November 2, 2015
04:56Laryngeal Mask Airway (LMA) Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway (SGA)
Published on: July 14, 2023
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...