Evaluation of mechanical heart valve size and function with ECG-gated 64-MDCT
Troy M LaBounty1, Prachi P Agarwal, Aamer Chughtai
1Department of Medicine, Division of Cardiology, Weill Cornell Medical College, New York, NY 10021, USA. tml9001@med.cornell.edu
Objective:
The purpose of our study was to determine whether CT can accurately evaluate mechanical heart valve size and function.
Materials And Methods:
Sixty-two patients with mechanical valves (37 single-disc, 27 bileaflet; 59 aortic, 5 mitral) were evaluated with ECG-gated 64-MDCT and transthoracic echocardiography; a subset of 10 patients underwent cinefluoroscopy. Two readers independently interpreted each study.
Results:
The mean age of the patients was 46.4 +/- 14.4 years; 50 were men and 12 were women. There was excellent correlation, and differences between CT readers were absent to small in measuring the opening angle (r = 0.96, p < 0.001; 76.7 +/- 9.0 degrees vs 76.8 +/- 9.6 degrees , p = 0.73), annulus diameter (r = 0.96, p < 0.001; 25.9 +/- 3.3 vs 25.9 +/- 3.2 mm, p = 0.62), and geometric orifice area (r = 0.98, p < 0.001; 3.8 +/- 0.9 vs 3.6 +/- 0.8 cm(2), p < 0.001). There was strong correlation without difference in opening angle between CT and cinefluoroscopy (r = 0.77, p < 0.001; 79.2 degrees +/- 9.8 degrees vs 77.2 degrees +/- 15.5 degrees , p = 0.45). Compared with manufacturer specifications, CT reported opening angles that were smaller for single-disc valves (n = 36, 67.4 degrees +/- 5.7 degrees vs 75 degrees , p < 0.001) and similar for bileaflet valves (n = 42 for 21 valves, 83.8 degrees +/- 3.9 degrees vs 85 degrees , p = 0.05), valves, with small underestimation with CT versus specifications in annulus diameter (n = 41; r = 0.75, p < 0.001; 26.4 +/- 3.0 vs 27.5 +/- 3.3 mm, p = 0.003), and geometric orifice area (n = 35; r = 0.90, p < 0.001; 3.7 +/- 0.7 vs 3.8 +/- 0.8 cm(2), p = 0.04). Each disc closed fully on CT; none had more than mild regurgitation on echocardiography.
Conclusion:
CT can measure the size and function of mechanical valves with high interobserver agreement and results similar to specifications. The opening angle with CT strongly correlates with cinefluoroscopy. CT is promising for the assessment of mechanical valves.
Related Concept Videos
Mitral Stenosis II: Clinical features and Diagnostic Tests
Imaging Studies for Cardiovascular System V: CT
Mitral Valve Prolapse II: Assessment and Management
Mitral Stenosis IV: Nursing Management
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...