Durability Analysis of a Self-Expanding Transcatheter Aortic Valve Through 30 Years of Accelerated Wear Testing
Guilherme Attizzani1,2, Marco Frazzetto1, Ole De Backer3,4
1Division of Cardiology, University Hospitals of Cleveland, Case Western Reserve University, Cleveland, Ohio, USA.
Structural Heart : the Journal of the Heart Team
|August 29, 2026
Abstract:
•First reported 30-year in vitro accelerated wear testing data for a THV.•All Evolut PRO + valves met ISO hydrodynamic criteria through 20 years of testing.•At least one valve per size maintained performance through 30 years of testing.•Failures manifested in a consistent pattern, suggesting a shared mechanism.•Testing is ongoing until all valves fail.