Imaging of cochlear implant electrode array with flat-detector CT and conventional multislice CT: comparison of image
Tobias Struffert1, Victoria Hertel, Yannis Kyriakou
1Department of Neuroradiology, Institute of Medical Physics, University of Erlangen-Nuremberg, Schwabachanlage 6, Erlangen, Germany. tobias.struffert@uk-erlangen.de
Conclusion:
Cochlear implantation assessment is possible using commercially available standard flat-detector computed tomography (FD-CT) protocols. Image quality is superior to multislice CT (MSCT). The radiation dose of FD-CT is lower in comparison with MSCT standard protocols and may therefore overcome the limitations of MSCT in the evaluation of cochlear implants.
Objective:
FD-CT offers higher spatial resolution than MSCT. Our objective was to compare the image quality of FD-CT to conventional MSCT in the visualization of a cochlear implant electrode array with respect to radiation exposure.
Methods:
An isolated temporal bone specimen was scanned using a commercially available FD-CT system and a 4 and 64 row MSCT scanner. Different scanning protocols were used. Image quality was assessed by four independent readers using a scoring system with different criteria describing delineation of the cochlea and the electrode array, image noise and spatial resolution. Radiation dose was measured using the CT dose index (CTDI) and a 16 cm acrylic phantom.
Results:
Image quality was rated superior for FD-CT for all criteria by all readers. Single electrode contacts were only visible in FD-CT and assessment of implant position was improved by FD-CT. The radiation dose of FD-CT was half that of MSCT standard protocols.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET