Quantitation of humalog insulin by reversed-phase high-performance liquid chromatography
Carey Munsick1, Robert Murray, Tom Dziubla
1Jefferson College of Graduate Studies and The Artificial Pancreas Center, Department of Anesthesiology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
Introduction:
The methodology for in vitro testing of insulin delivery systems for long-term infusion of lispro insulin requires insulin flow studies over time, with the measurement of lispro concentrations in wells or other laboratory fluid collection systems. We postulated that the efficiency of the insulin assay could be improved if the insulin collected in the wells could be measured without having to perform dilutions of insulin samples. The manufacturer's method for the identification of Humalog(R) insulin by high-performance liquid chromatography (HPLC) does not provide a detailed method for the quantitation of the clinical formulation. No reports could be found in the literature describing a method to quantitate lispro insulin's entire concentration range, nor could we find detailed information on pH effects. The purpose of this study was to investigate the use of a reversed-phase HPLC method to quantitate Humalog insulin at three levels of pH.
Methods:
Serial dilutions of Humalog insulin stock solution were prepared at pH levels of 2.6 and 7.4 (concentrations 0.20-100 IU/ml) to construct calibration curves. Samples were also prepared at lower concentrations (0.23-15.0 IU/ml) at pH levels of 4.0 and 7.4. Areas under curve data were plotted against known concentrations, and the limit of quantitation and precision of the assay were determined.
Results:
Calibration curves for the 0.20-100 IU/ml concentration (pH 7.4 and 2.6) revealed no significant differences (p < 0.01) between those curves. pH and concentration did not have an effect on the overall precision of the method. Linearity was not conserved past the 15-IU/ml concentration. Coefficient of variation (CV) values were generally <15% except at the lower concentrations where the largest CVs were observed. Mean known concentrations plotted on a logarithmic scale demonstrated linearity from 12.5 to 100 IU/ml.
Conclusions:
Humalog insulin clinical formulation can be quantitated over the entire concentration range by a reversed-phase HPLC method using nonlinear regression analysis. The method is reproducible at lower (<15 IU/ml) and higher insulin concentrations. Linear regression analysis may be used when the concentrations of interest are in the 0- to 15-IU range. Preparation of insulin solutions at pH 2.6, 4.0, and 7.4 did not significantly affect the reproducibility of the assay.
More Related Videos
10:21Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification
Published on: September 21, 2011
10:52Radiolabeling and Quantification of Cellular Levels of Phosphoinositides by High Performance Liquid Chromatography-coupled Flow Scintillation
Published on: January 6, 2016
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors