Laboratory-based non-clinical comparison of occlusion rates using three rapid-acting insulin analogs in continuous
David Kerr1, John Morton, Caroline Whately-Smith
1Bournemouth Diabetes and Endocrine Centre, Royal Bournemouth Hospital NHS Foundation Trust, Dorset, UK. david.kerr@rbch.nhs.uk
Background:
Rapid-acting analog insulin is used increasingly for continuous subcutaneous insulin infusion therapy (CSII). As the choice of insulin may be a determinant of catheter occlusion, we compared rates of early and late occlusion of a standard CSII catheter with three insulin analogs in a laboratory-based setting.
Methods:
Twenty-four pumps were used for the study. Each insulin analog (glulisine, lispro, and aspart) was assigned to eight pumps in a randomized order for each of nine runs of 5-day duration. Pumps were primed to receive a basal dose of 0.1 IU/h with a bolus dose of 2 IU given three times each day. Pumps were placed in an incubator to maintain temperature in the range of 32 to 36 degrees C.
Results:
Over the entire study period, there were 48 occlusions. Early occlusions (within 72 hours) occurred during five of the nine runs with no evidence of any difference between insulins (p = .27); there were no occlusions before 48 hours. Over the whole of the 5-day infusion period, the probabilities of overall occlusion for each insulin were 40.9% [28 to 55%, 95% confidence interval (CI)] for glulisine, 9.2% (4 to 19.5%, 95% CI) for aspart, and 15.7% (8.1 to 28.1%, 95% CI) for lispro. All occlusions, except for three, occurred during a bolus infusion.
Conclusions:
During CSII under laboratory conditions, early catheter occlusions (within 72 hours) are rare and independent of the choice of insulin analog. For patients using insulin pump therapy, the importance of catheter change within 72 hours should be emphasized irrespective of the insulin used. Beyond 72 hours, the risk of occlusion differs between insulins, being more common with glulisine.
Related Concept Videos
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into rapid-acting...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...