Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer
Published on: February 20, 2020
Comparison of hand dynamometers in elderly people
1Faculdade de Ciências da Nutrição e Alimentação da Universidade do Porto, Portugal.
The Journal of Nutrition, Health & Aging
|November 20, 2009
Summary
Hand grip strength dynamometers showed excellent laboratory accuracy but varied performance in the elderly. The Smedlay
Area of Science:
- Gerontology
- Biomedical Engineering
- Rehabilitation Science
Background:
- Previous dynamometer assessments in adults highlighted issues with practicality, weight, and ergonomics.
- More suitable tools are needed for gerontological and clinical research involving older adults.
Purpose of the Study:
- To evaluate the measurement reliability and performance of various hand grip strength dynamometers in an elderly population.
- To compare the accuracy and agreement of four static grip strength dynamometers against a reference standard in older adults.
Main Methods:
- A cross-sectional study was conducted in Porto, Portugal, involving a retirement home and a social day care center.
- Fifty-five elderly individuals (65-99 years) had their grip strength measured using four dynamometers: Smedlay's Hand, Sammons Preston Rolyan Bulb, Eisenhut, and Jamar Hydraulic Hand.
- The Jamar Hydraulic Hand served as the comparison dynamometer.
Main Results:
- Laboratory accuracy of all four dynamometers against known forces was excellent (r > 0.96).
- A strong association was found between the Jamar Hydraulic and the other instruments (r > 0.77), but significant differences in mean grip strength values were observed.
- Bland-Altman plots indicated poor agreement between the Jamar Hydraulic and the other three dynamometers.
Conclusions:
- While all four dynamometers demonstrated high accuracy in laboratory settings, their performance varied significantly when used with elderly individuals.
- The Smedlay's dynamometer showed results closest to the Jamar Hydraulic, but none of the tested dynamometers provided comparable measurements to the Jamar Hydraulic in this population.