Mucosal and serosal changes after gastric stapling determined by a new "real-time" surface tissue oxygenation probe:

Christopher J Myers1, Gevorg Mutafyan, Aurora D Pryor

  • 1Department of Surgery, Duke University Medical Center, Durham, North Carolina 27710, USA. chrismyers2@hotmail.com

Summary

A new tissue oximeter probe accurately measured gastric tissue oxygen saturation (StO(2)) during stapling. Significant StO(2) drops occurred near staple lines, even without visible changes, indicating potential risks for surgical complications.