Comparison of polyp size and volume at CT colonography: implications for follow-up CT colonography
Emily Bethea1, Ogonna K Nwawka, Abraham H Dachman
1Department of Radiology, Pritzker School of Medicine, The University of Chicago, MC 2026, Chicago, IL 60637, USA.
Objective:
The purpose of this study was to evaluate the reliability of polyp measurements at CT colonography and the factors that affect the measurements.
Materials And Methods:
Fifty colonoscopically proven cases of polyps 6 mm in diameter or larger were analyzed by two observers who measured each polyp in supine and prone views. Manual measurements of 2D volume by summation of areas, 2D maximum diameter, and 3D maximum diameter and automated measurements of 3D maximum diameter and volume were recorded for each observer and were repeated for one of the observers. Intraobserver and interobserver agreement was calculated. Analysis was performed to determine the measurement parameter that correlated most with summation-of-areas volume. Supine and prone measurements as a surrogate for tracking change in polyp size over time were analyzed to determine the measurement parameter with the least variation.
Results:
Maximum diameter measured manually on 3D images had the highest correlation with summation-of-areas volume. Manual summation-of-areas volume was found to have the least variation between supine and prone measurements.
Conclusion:
Linear polyp measurement in the 3D endoluminal view appears to be the most reliable parameter for use in the decision to excise a polyp according to current guidelines. In our study, manual calculation of volume with summation of areas was found to be the most reliable measurement parameter for observing polyp growth over serial examinations. High reliability of polyp measurements is essential for adequate assessment of change in polyp size over serial examinations because many patients with intermediate-size polyps are expected to choose surveillance.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Endoscopic Procedures II: Colonoscopy
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies VI: Voiding Cystourethrography and Cystography
Imaging Studies for Cardiovascular System V: CT