Simple test pattern for mammographic screen-film contact measurement
1Department of Radiology, University of Tennessee, Memphis 38168.
Radiology
|March 1, 1991
Summary
Radiographic grids can test mammographic screen-film contact, eliminating the need for specialized fine-mesh patterns. Magnification may be necessary due to high grid line frequencies in modern mammography.
Area of Science:
- Radiology
- Medical Imaging Physics
Background:
- Mammographic screen-film contact is crucial for image quality.
- Specialized fine-mesh test patterns are typically required for mammography.
- Existing test patterns for other diagnostic cassettes use coarser meshes.
Purpose of the Study:
- To identify alternative, readily available test patterns for assessing mammographic screen-film contact.
- To determine if standard radiographic grids can be used for mammography testing.
Main Methods:
- Investigated the use of radiographic grids with varying line spacings.
- Evaluated grids with line spacing equal to or exceeding 40 lines per inch.
- Considered the application of magnification techniques for testing.
Main Results:
- Radiographic grids with line spacing of 40 lines per inch or greater are effective for testing mammographic screen-film contact.
- This approach negates the need for purchasing dedicated fine-mesh mammographic test patterns.
- Magnification techniques are often required due to modern grids exceeding 100 lines per inch.
Conclusions:
- Standard radiographic grids offer a cost-effective alternative for mammographic screen-film contact testing.
- Healthcare facilities can utilize existing grid resources, improving accessibility and reducing costs.
- Proper technique, including magnification, ensures accurate assessment with these alternative patterns.