Prashant N Chhajed1, David Miedinger, Florent Baty
1Clinic for Pulmonary Medicine, University Hospital, Basel, Switzerland.
Background:
Cutaneous carbon dioxide tension (PcCO(2)) is a promising non-invasive surrogate measure of arterial partial pressure of carbon dioxide (PaCO(2)).
Objectives:
To compare values of PcCO(2) and oxygen saturation (SpO(2)) with arterial blood gas (ABG) analysis.
Methods:
SpO(2) and PcCO(2) were measured with a v-Sign-sensor (Sentec AG, Therwil, Switzerland) and the values compared with simultaneously obtained SaO(2) and PaCO(2) obtained from ABG analysis (ABL 725, Radiometer, Copenhagen, Denmark) in 275 adult patients referred to the lung function laboratory.
Results:
Median of the PcCO(2) was 4.7 kPa (interquartile range [IQR] 0.9 kPa). Median of the SpO(2) was 97% (IQR 3%). Bland-Altman analysis for comparison of PcCO(2) with PaCO(2) showed a bias of -0.1 kPa with a precision of +/- 0.9 kPa with 3.7% outlying values. Bland-Altman analysis for the comparison of SpO(2) and SaO(2) showed a bias of 20.1 % with a precision of +/- 3.5%. There were no complications.
Conclusion:
There is a good agreement between combined cutaneous capnography and oximetry values with ABG analysis. Due to the excellent safety profile and the short time to get a continuous measurement, this technique should be examined in settings where it can complement repeated ABG analysis when ventilatory disturbances are suspected or non-invasive monitoring of ventilation is needed.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and HCO3− values are examined to...