A protocol design for evaluation of wearable cuff-less blood pressure measuring devices
Iris R F Yan1, Carmen C Y Poon, Y T Zhang
1Joint Research Center for Biomedical Engineering (JCBME), Dept. of Electronic Engineering, The Chinese University of Hong Kong, Hong Kong, China.
Abstract:
This paper proposes one of the unique requirements in the validation protocol of the IEEE P1708 standard in assessing wearable cuff-less blood pressure (BP) measuring devices. Based on principles that are different from that of the conventional cuff-based devices, the cuff-less BP measurement approaches often require an individual calibration procedure. In this study, we used data from an experiment carried out on 28 subjects with a total of 139 sets of BP measurements as an example to show that breakdown of the performance evaluation of cuff-less devices according to the change of BP from the point of calibration is crucial for understanding and interpreting the overall accuracy of the device.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)