Evaluation of a fiber-optic esophageal pulse oximeter
Justin P Phillips1, Richard M Langford, Serene H Chang
1School of Engineering and Mathematical Sciences, City University, London, EC1V 0HB, UK. justin.phillips.1@city.ac.uk
Abstract:
A dual-wavelength fiber-optic pulse oximetry system was developed for the purposes of estimating oxygen saturation from the esophagus. A probe containing miniature right-angled glass prisms was used to record photoplethysmographic (PPG) signals from the esophageal wall. Signals were recorded successfully in 19 of 20 patients, demonstrating that PPG signals could be reliably obtained from an internal vascularized tissue site such as the esophageal epithelium. The value of the mean oxygen saturation recorded from the esophagus was 94.0 +/- 4.0%. These results demonstrate that SpO(2) may be estimated in the esophagus using a fiber-optic probe and this may be the first report of such measurements.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...