Calibration of cross-sectional images measured by an ultrasound-based muscle evaluation system
Kiyotaka Fukumoto1, Satoshi Muraki, Masayoshi Tsubai
1Graduate School of Design, Kyushu University, 4-9-1, Shiobaru, Minami-ku, Fukuoka, Fukuoka, 815-8540, Japan.
Abstract:
The quantification of muscle volume can be used to estimate muscular strength. Therefore, we developed a flexible measuring system for muscle volume using ultrasonog-raphy. In the measuring process, subjects are not required to perform any muscular contraction, so it is completely safe and particularly suitable for elderly people. The ultrasound probe is installed on a mechanical arm, and continuously scans fragmental images along the body surface. The measured images are then composed into a wide area cross-sectional image. However, the muscle area measured by our system was slightly smaller than that measured by MRI. because the ultrasound probe contacted the body surface with a little pressure during the measurement. The strain then decreases the total image size and its circumference. This paper introduces our developed system and proposes a new calibration method for the muscle area in the thigh based on its circumference.