Related Experiment Video
Updated: Jun 18, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Verification of air-kerma strength of 125I seed for permanent prostate implants in Japan
Iori Sumida1, Masahiko Koizumi, Yutaka Takahashi
1Department of Radiation Oncology, Osaka University Graduate School of Medicine, 2-2 Yamada-oka, Suita, Osaka 565-0871, Japan. sumida@radonc.med.osaka-u.ac.jp
Background:
To assure the physical quality of brachytherapy, we investigated the difference between measured and manufacturer's stated source strengths in a single model SourceTech Medical (STM)1251 (125)I seed.
Methods:
A well-type ionization chamber with a single-seed holder was used to measure the source strength of 2412 (125)I seeds before implant in 34 patients. The air-kerma strength was 0.450 U for all cases. The mean source strength for each patient was measured and compared with the manufacturer's stated value. The deviation from the measured value was compared with the tolerance range of the American Association of Physicists in Medicine (AAPM) TG-56 report's recommendation.
Results:
The measured source strength was higher than the manufacturer's stated value, with a median difference of 1% (range, 2% to 5%). Sixteen of the total of 2412 seeds (0.7%) were more than 5% different from the manufacturer's stated value. The median SD from the mean value was 2.2% (range, 1.1% to 2.5%) for all patients.
Conclusion:
This is the first report of a single-seed assay performed for the model STM1251 (125)I seed. In this study the manufacturer's stated strength agreed well with the measured value. Nevertheless, the advisability of performing a single-seed assay at every institution should be considered, by referring to the appropriate regulations; for example, those used in the United States.