Related Experiment Video
Updated: Jun 17, 2026

Microfluidic Platform with Multiplexed Electronic Detection for Spatial Tracking of Particles
Published on: March 13, 2017
First results in the development of an on-line digital counting platform dedicated to primary measurements
C Bobin1, J Bouchard, B Censier
1CEA, LIST, Laboratoire National Henri Becquerel (LNE-LNHB), Gif-sur-Yvette, France. christophe.bobin@cea.fr
Abstract:
This paper describes the first step in the development of a digital platform dedicated to primary radionuclide measurements. In order to validate the implementation of an on-line processing based on extendable dead-times, a FPGA-based digital system has been programmed for liquid scintillation counting. After a description of the hardware, TDCR (Triple to Double Coincidence Ratio) measurements are presented. Comparisons of activity measurements with the classical system operating at LNHB are used to validate the new system; coincidence counting differences observed when increasing the resolving time in the digital platform are discussed.
Related Concept Videos
Electronic Distance Measuring Instruments
Measurement: Derived Units
Measurement: Standard Units
Microbial Growth Measurement: Direct Methods
Coulometry: Overview
Measurement of Blood Pressure